Helios Focused Ion Beam (FIB) Laboratory
The instrument is an FEI Helios NanoLab 600i DualBeam FIB/SEM, containing both a focused Ga+ ion beam ("Tomahawk") and a high resolution field emission scanning electron ("Elstar") column. Combined with advances in patterning, scripting, and a suite of accessories, these features make milling, imaging, analysis, and sample preparation down to the nanoscale possible as standard applications in the lab.
Specifications:
- Elstar UHR immersion lens FESEM column
- Accelerating voltage range: 0.35 – 30 kV
- Beam deceleration with stage bias from -50 V to -4 kV
- Landing voltage range 20 V – 30 kV
- Probe current range: 0.7 pA to 22 nA
- Electron beam resolution at optimum WD
- 0.8 nm at 30 kV (STEM)
- 0.9 nm at 15 kV
- 1.4 nm at 1 kV
- Electron beam resolution at coincident point
- 0.9 nm at 15 kV
- 1.6 nm at 5 kV
- 2.5 nm at 1 kV
- Tomahawk ion column
- Accelerating voltage range: 0.5 – 30 kV
- Probe current range: 0.1 pA – 65 nA with 15 apertures
- Time-of-flight (TOF) correction
- Ion beam resolution at coincident point
- 4.0 nm at 30 kV using preferred statistical method
- 2.5 nm at 30 kV using selective edge method
- High precision 5-axes motorized stage
- XY: 150 mm, piezo-driven
- Z: 10 mm, motorized
- T: -10º – 60º
- R: n x 360º (endless), piezo-driven
- Tilt accuracy (between 50º to 54º): 0.1º
- X, Y repeatability: 1.0 µm
- Detectors
- Elstar in-lens SE detector (TLD-SE)
- Elstar in-lens BSE detector (TLD-BSE)
- Everhart-Thornley SE detector (ETD)
- IR camera for viewing sample/column
- Door-mounted Nav-Cam
- High performance Ion Conversion and Electron (ICE) detector for secondary ions (SI) and electrons (SE)
- Retractable low voltage, high contrast solid-state electron detector (DBS)
- Retractable STEM detector with BF/DF/HAADF segments
- Beam current measurement
- Chamber
- 4mm E- and I-beam coincidence point at analytical WD
- Angle between electron and ion columns: 52º
- Maximum sample size: 150 mm diameter with full rotation
- Maximum clearance between stage and coincidence point: 55 mm
- Sample Weight: maximum 500 g (including the sample holder)
- Image processor
- Dwell time range from 0.025 to 25,000 µs/pixel
- up to 6144 x 4096 pixels
- SmartSCAN (256 frame average or integration, line integration and averaging, interlaced scanning) and DCFI (Drift Compensated Frame Integration)
- Supporting software
- "Beam per quad" graphical user interface concept, with up to 4 simultaneously active quads
- FEI SPI, iSPI, iRTM and FIB immersion mode for advanced, real-time SEM And FIB process monitoring and endpointing
- Patterns supported: lines, rectangles, polygons, circles, annuli, cross-sections, and cleaning cross-sections
- Image registration
- Directly imported BMP file or streamfile for 3D milling and deposition
- Material file support for "minimum loop time", beam tuning and independent overlaps
- Accessories
- Gas Injection Systems
- MultiChem (TM)
- Tungsten deposition
- Carbon deposition
- Insulator enhanced etch (XeF2)
- Selective Carbon Mill (patented)
- Platinum deposition
- Omniprobe AutoProbe 200 in situ sample lift-out system
- Quorum PP3000T cryo system
- FIB Charge Neutralizer
- Ingerated Plasma Cleaner
- FEI CryoCleaner
- AutoFIB
- macro and script-based DualBeam automation
- MAPS
- automatic acquisition of large images and correlative work
- AutoTEM
- automated TEM sample preparation with sectioning
- GDStoDB and NanoBuilder
- basic and advanced FEI proprietary CAD-based (GDSII) solutions for FIB and beam deposition optimized nanoprototyping of complex structures
- AutoSlice and View
- automated sequential mill and view to collect series of slice images for 3D reconstruction
Restrictions on samples
The sample material must be able to withstand a high vacuum environment without outgassing. It must be clean. It may be attached to the sample holder using a suitable SEM vacuum-quality adhesive (please check with us first). The sample must be electrically grounded to the sample holder to minimize sample charging. If the sample is nonconductive (plastic, fiber, polymer, or other substance with an electrical resistance greater than 1E10 ohms), certain strategies may need to be employed for successful work, and please check with us regarding those.
Training and Service
Before being considered for training on the FEI Helios NanoLab FIB/SEM, you must first be a fully qualified user of the FEI DB235 Dual-Beam FIB/SEM. Further information is on the training info page.
Service requests will be considered on a case-by basis. However, our normal mode of operation is to train users to perform the characterization experiments themselves.
(information from FEI)