Research Topics
Experimental Methods
|
Ultrafast 2D IR Vibrational Echo Spectroscopy

Figure 19 is a schematic representation of the experiment and the pulse
sequence. There are three excitation pulses, ~60 fs in duration produced
with a Ti:Sapphire regenerative amplifier pumped optical parametric
amplifier (OPA) (see below). The pulse sequence is shown at the bottom of
the figure. The OPA is tuned to the frequency of the vibrational mode under
investigation. The IR pulses have sufficient bandwidth to span the spectral
region of interest. The times between pulses 1 and 2 and between pulses 2
and 3 are called t, and Tw, respectively. The vibrational echo signal
radiates from the sample at a time £t after the third pulse in a unique
direction. The vibrational echo signals are recorded by scanning t at fixed
Tw. The signal is spatially and temporally overlapped with a local
oscillator for heterodyne detection, and the combined pulse is dispersed by
a monochromator onto an IR array detector. Heterodyne detection provides
both amplitude and phase information. If the vibrational echo is detected as
it emerges from the sample, there is no phase information. The experiment
produces a 2D spectrum in the frequency domain. The measurements are made in
the time domain. To go from the time domain to the frequency domain requires
two Fourier transforms. To perform a Fourier transform requires both
amplitude and phase information. The local oscillator is another IR fs
pulse. When it overlaps with the vibrational echo wave packet, interference
occurs. As shown below, when the time t is scanned, the vibrational echo
pulses moves in time across the fixed local oscillator pulse. The result is
a temporal interferogram that provides the necessary phase information.
As shown in figure 19, the combined vibrational echo/local oscillator pulse
is passed through a spectrometer and recorded by an IR array detector. The
array detector allows us to measure 32 wavelengths simultaneously, reducing
the data acquisition time by a factor of 32. Taking the spectrum of the
heterodyne-detected vibrational echo signal performs one of the two Fourier
transforms and gives the wm axis (m for monochromator) in the 2D IR spectra.
When t is scanned, a temporal interferogram is obtained at each wm. The
temporal interferograms are numerically Fourier transformed to give the
other axis, the wt axis. 2D IR spectra are obtained for a range of Tws.

Figure 20 shows a very simple example of the nature of the data. The upper
left portion of the figure shows the spectrum of the hydroxyl stretching
mode of phenol-OD. The H in the OH hydroxyl group has been replaced by a D.
The IR absorption spectrum shows a single peak. The structure of the
molecule is shown as an inset. As outlined to the right of the spectrum,
when t is scanned, the time of the output of the vibrational echo electric
field, which is the signal (S), changes relative to the fixed in time of the
local oscillator electric field (L). L is much larger than S. The detector
measures the absolute value squared of the sum of the electric fields. This
gives the three terms. L2 is constant in time. S2 is very small and not
detected relative to the much larger cross term, 2LS(t). As t is scanned,
the temporal interferogram is created. A portion of an interferogram is
shown in the bottom left side of the figure. There is one such interferogram
for each monochromator frequency, wm. wm is the vertical axis of the 2D IR
spectrum shown in the lower right portion of the figure. This axis is
obtained from the frequency measurements performed by the monochromator. The
horizontal axis, the wt axis, is obtained by numerical Fourier
transformation of interferograms like that shown in the figure. There is one
interferogram at each wm where there is signal.
The 2D IR vibrational spectrum at the time Tw = 16 ps has two peaks. The
peak on the diagonal (the dashed line) is positive going (red). The peak
off-diagonal is negative going (blue). The wt axis is the frequency of the
first interaction of the molecules with the radiation field (first pulse).
This first interaction is represented by the dashed arrow on the left side
of the energy level diagram that is to the right of the spectrum in figure
20. The first pulse makes a coherent superposition state of the vibrational
ground state (0) and the first vibrationally excited state, (1). The second
pulse, represented by the solid arrow in the energy level diagram transfers
the vibrational coherence to populations in the 0 and 1 levels. Phase
information is stored in the populations. The third pulse between 0 and 1
(dashed arrow between 0 and 1) again produces a coherent superposition state
(a coherence) which gives rise to the vibrational echo emission (wavy arrow)
at the frequency the molecules interact with the third pulse. The wm axis is
the axis of the vibrational echo emission. The 0-1 vibrational peak is on
the diagonal because the first interaction of the molecules with the first
radiation field (wt axis) is at the same frequency as the last interaction
and vibrational echo emission (wm axis).

The off-diagonal 1-2 peak in the spectrum is shifted along the wm axis by
the vibrational anharmonicity (unequal spacing of the energy levels as shown
in the energy level diagram). The first two interactions are the same as
discussed above. They produce population in the 1 level. Because the
bandwidth of the laser is large, it is greater than the anharmonic shift of
the levels. So the third interaction can produce a coherence between the 1
and 2 levels, which is represented by the dashed arrow connecting the 1 and
2 levels. The 1-2 coherence gives rise to vibrational echo emission at the
1-2 transition frequency. Because the first interaction (wt axis) is at the
0-1 transition frequency and the third interaction and vibrational echo
emission (wm axis) is at the shifted 1-2 transition frequency, the 1-2 peak
appears off-diagonal. The difference between diagonal and off-diagonal peaks
is very important. If the first and last interactions are at the same
frequency, a peak will be on the diagonal. If the first and last
interactions are at different frequencies, a peak will be off-diagonal.
Figure 21 is a simplified schematic of the equipment used to perform the 2D
IR vibrational echo experiments. The upper portion of the figure shows the
laser equipment. A CW diode pumped neodymium vanadate laser is used to pump
a Ti:Sapphire oscillator, which seeds a Ti:Sapphire regenerative amplifier.
The regen is pumped with 7 watts by a 1 kHz diode pumped Nd:YLF Q-switched
laser. The output of the regen at 800 nm is 2/3 mJ, 35 fs pulses. The 800 nm
pulses drive a multi-stage OPA, which produces 3 uJ pulses with 60 fs
duration. These are extremely short pulses. At a wavelength of 4 mm, the
pulses are only ~4 cycles of light. This number of cycles of light would be
the equivalent of an ~6 fs pulse with a wavelength in the middle of the
visible spectrum. The lower portion of figure 21 shows the optical setup
that does the vibrational echo experiment. The input IR pulse from the OPA
is split into 5 beams using beam splitters (BS). Three of them are the three
excitation pulses, 1, 2, and 3 in the pulse sequence shown at the bottom of
figure 19. A fourth pulse is the local oscillator (see figure 19). The fifth
pulse is called the tracer. It is aligned along the path that the
vibrational echo will emerge from the sample (see figure 19) and is used
only for alignment purposes. It is blocked during the experiments. Pulses 1,
2, and the LO are passed down ultra-precision translation stages labeled 1,
2, and LO. These stages can take 10 nm steps and are exceedingly
reproducible in their position.
The ultra-precision is necessary because the
vibrational echo is a type of IR holography. The interferogram shown in
figure 20 must be accurate and reproducible. The translation stages labeled
3 and tr (tracer) are manual translation stages. Once these timings are set,
they do not change. The three excitation beams are focused into the sample
using an off-axis parabolic reflector (P1). After passing through the sample
the three excitation beams and the vibrational echo beam are collimated by a
second off-axis parabolic reflector (P2). The beams other than the
vibrational echo are blocked. The LO is split into two beams. One is
combined with the vibrational echo. The combined pulses pass through the
monochromator and are detected by the upper stripe of the dual stripe IR 32
element array detector. The other portion of the LO passes through the
monochromator and is detected by the lower stripe of the array. The signal
from the lower stripe is used to normalize the signal on the upper stripe
for the IR spectrum and shot-to-shot intensity fluctuations. The signals on
the array detector are read out after each laser shot by a computer for
processing into the 2D IR vibrational echo spectrum.
Ultrafast 2D IR Spectroscopy Recent Publications
311. “Vibrational Echo Correlation Spectroscopy Probes of Hydrogen Bond
Dynamics in Water and Methanol,” John B. Asbury, Tobias Steinel, and M. D.
Fayer, J. Lumin. 107, 271-286 (2004).
332. “Ultrafast Dynamics of Solute-Solvent Complexation Observed at Thermal
Equilibrium in Real Time,” Junrong Zheng, Kyungwon Kwak, John Asbury, Xin
Chen, I. Piletic, and M. D. Fayer, Science 309, 1338-1343 (2005).
361. “Ultrafast 2D-IR Vibrational Echo Spectroscopy: A Probe of Molecular
Dynamics,” Sungnam Park, Kyungwon Kwak, and M. D. Fayer Laser Phys. Lett. 4,
704-718 (2007).
363. “Frequency-Frequency Correlation Functions and Apodization in 2D-IR
Vibrational Echo Spectroscopy, a New Approach,” Kyungwon Kwak, Sungnam Park,
Ilya J. Finkelstein, and M. D. Fayer, J. Chem. Phys. 127, 124503 (17 pages)
(2007).
374. “Taking Apart 2D-IR Vibrational Echo Spectra: More Information and
Elimination of Distortions,” Kyungwon Kwak, Daniel E. Rosenfeld, and M. D.
Fayer J. Chem. Phys. 128, 204505 (2008).
>>
Next: Polarization Selective Pump-Probe [ Top
|
Previous: Complex Liquids ]
[ Home | Research
Overview | Group Members
| Quantum Mechanics Book |
Publications ] |